CAN CELL VIABILITY REAGENTS DAMAGE YOUR VI-CELL?
The Vi-Cell Pump assembly, manufactured by Klohen (Kloehn Norgren, part of IMI Precision Engineering), incorporates the syringe pump assembly along with the distribution valve for fluidics. This pump interfaces bi-directionally through a 9-pin serial cable and COM port on the computer. It sends commands to the various Vi-CELL electrical components via the Control PCB and contains two stepper motors: one rotates the distribution valve to select the proper fluidics from the reagent pack, and one controls the syringe operation for aspiration of fluids. The Pump also contains the CPU function, the NOVRAM where the calibration data, focus data, serial number, and firmware and various other memory values are stored.
During operation, the Vi-Cell Syringe pump aspirates cell cultures and the individual reagent components (Trypan Blue, Water, Enzymatic Cleaner and Disinfectant) via thin PTFE tubing into the glass syringe. The glass syringe is also responsible for pushing the sample to the Quartz Flow Cell via similar tubing and later does a cleaning in preparation for the next sample. The figure below shows the basic fluidic layout of the Vi-Cell system.
Parts that are in contact with fluids from the reagent pack, from Beckman Coulter or Cell Viability Inc, are as follows:
Teflon (PTFE) Tubing Quartz Flow cell
Distribution valve with fittings and Teflon washers Glass Syringe
Sipper arm (PTFE and Stainless Steel) Waste bottle
ALL parts listed above are chemically resistant to reagent pack component fluids found in both the Beckman Coulter and Cell Viability reagent packs.
With respect to equivalency, below is a list of the components of both reagent packs. The table below was created using the Material Data Safety Sheets for both the Beckman Coulter and Cell Viability reagent packs.
Component | Beckman Coulter Name & Contents |
Cell Viability Name & Contents | Equivalency |
Trypan Blue Dye | Trypan Blue 0.4% | Trypan Blue Solution, 0.4% | Equivalent |
Disinfectant | Disinfectant: Isopropyl Alcohol, % by wt., 70 |
Isopropyl alcohol, 70% in water | Equivalent |
Buffer Solution | Buffer Solution | Buffer Solution | Equivalent |
Cleaning Agent | Cleaning Agent | Cleaning Agent- Enzymatic Cleaner | Equivalent |
Showing the above, it is evident that the two reagent packs are fully equivalent and under normal circumstances, neither can damage the Vi-Cell system.
Please be sure to see additional Technical Note on decontamination process required to remove excessive sample protein buildup
For additional Information, please contact us directly:
Cell Viability Inc
14359 Miramar Parkway, #495
Miramar Florida 33027
Phone – 1-800-618-7660
E-mail: Sales@CellViability.com
www.CellViability.com
The information listed on this Question and Answer section has material presented by Cell Viability Incorporated which is intended to be advice, opinions and for informational purposes only. This website and information is not intended to constitute official technical advice nor the provision of official technical services. By posting and/or maintaining this website and its contents, Cell Viability Incorporated does not intend to imply official technical information for instrumentation belonging to other companies.
Teflon is a registered trademark of Chemours