HOW DO YOU DECONTAMINATE A VI-CELL

HOW DO YOU DECONTAMINATE A VI-CELL

There are two ways to decontaminate the Vi-CELL instrument: internally and externally.

 

External decontamination:

 

Remove all dried blood or cell culture media from the instrument surface before performing the internal decontamination.

 

1.-To remove these substances and prevent scattering potentially biohazardous material, the blood or culture media should be wetted and softened with the 0.5% sodium hypochlorite solution.

 

2.- After removal of the residual material, decontaminate the surface of the Vi-CELL instrument with the bleach solution again.

 

3.- If complete removal is not possible after those two attempts, apply 10% bleach solution to the instrument surface for 20-30 minutes. Then, remove the residual material.

 

4.- Rinse the surface with water to remove the bleach.

Internal decontamination

 

When a Flow Cell becomes dirty on the inside due to sample build-up, it often shows up an image as streaks across the screen.

Generally, this situation can be corrected by performing a system decontamination.

 

In the Vi-CELL software, there is a Decontamination Wizard which assist the operator through this standard procedure for the internal decontamination.

 

______________________________________________________________________

 

WARNING

 

Appropriate safety equipment must be worn: eye protection, latex gloves, and lab coat.

______________________________________________________________________

 

 

To execute the system decontamination:

 

1.- Under toolbar select Instrument.

 

2.- Then, from the dropdown box select Decontaminate and follow the prompts.

 

Depending on every case, repeat the routine as needed.

 

 

 

 

.

 

 

 

 

 

 

Cell Viability Inc
14359 Miramar Parkway, #495
Miramar Florida 33027
Phone – 1-800-618-7660
E-mail: Sales@CellViability.com
www.CellViability.com

 

 

The information listed on this Question and Answer section has material presented by Cell Viability Incorporated which is intended to be advice, opinions and for informational purposes only. This website and information is not intended to constitute official technical advice nor the provision of official technical services. By posting and/or maintaining this website and its contents, Cell Viability Incorporated does not intend to imply official technical information for instrumentation belonging to other companies.