WHAT TO DO WHEN YOU NEED TO SHUT DOWN YOUR VI-CELL FOR LONG TERM

WHAT TO DO WHEN YOU NEED TO SHUT DOWN YOUR VI-CELL FOR LONG TERM

The Vi-CELL instrument does not require a daily shutdown. However, it is highly recommended to perform a shutdown routine when you planning to halt your own process temporarily or for a period more than two week.

 

To complete a shutdown routine, perform the following steps:

 

______________________________________________________________________

 

WARNING

 

Appropriate safety equipment must be worn: eye protection, latex gloves, and lab coat.

______________________________________________________________________

 

 

  1. Under toolbar select Instrument.

 

  1. Then, from the dropdown box select Decontaminate and follow the prompts.

 

  1. Once completed, perform a Drain system.

Cell Viability Inc
14359 Miramar Parkway, #495
Miramar Florida 33027
Phone – 1-800-618-7660
E-mail: Sales@CellViability.com
www.CellViability.com

 

 

The information listed on this Question and Answer section has material presented by Cell Viability Incorporated which is intended to be advice, opinions and for informational purposes only. This website and information is not intended to constitute official technical advice nor the provision of official technical services. By posting and/or maintaining this website and its contents, Cell Viability Incorporated does not intend to imply official technical information for instrumentation belonging to other companies.